Technical Specifications
- Dimensions BiPAP (H x W x D): 4.6” x 6.5” x 5.7” @ 2.9 lbs
- Heated Humidifier (H4i) (H x W x D): 5.0” x 6.7” x 8.3” @ 1.6 lbs
- Symbols which appear on the device: Attention, consult accompanying documents; Drip proof; Type CF equipment; Dangerous voltage; Class II equipment; Start/Stop; Manufacturer.
- Operating pressure range: 3 to 25 cm H2O
- Maximum single fault steady state pressure: 40 cm H2O
- Pressure measurement tolerance: ±0.5 cm H2O ± 4% of the measured reading
- Flow measurement tolerance: ±0.1 or 20% of reading, whichever is greater
- VAuto mode: 4 to 25 cm H2O (measured at the mask); Min EPAP: 4 cm H2O; Max IPAP: 25 cm H2O; Pressure Support: 0 to 10 cm H2O S, ST and T modes IPAP: 4 to 25 cm H2O (measured at the mask); EPAP: 3 to 25 cm H2O (measured at the mask);
- Pressure Support: 0 to 22 cm H2O
- CPAP mode: 4 to 20 cm H2O (measured at the mask)
- DECLARED DUAL-NUMBER NOISE EMISSION VALUES in accordance with ISO 4871
- Sound pressure level: 26 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1: 2002
- 28 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
- Sound power level: 36 dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2007
- Dimensions (L x W x H): 4.4” x 6.5” x 5.7” (112 mm x 164 mm x 145 mm)
- Weight: 2.9 lbs (1.3 kg)
- Power supply Input range :100–240 V, 50–60 Hz, 40 VA (typical power consumption), < 100 VA (maximum power consumption)
- Housing construction: Flame retardant engineering thermoplastic
- Operating temperature: 41ºF to 95ºF ( 5ºC to 35ºC)
- Operating humidity: 10–95% non-condensing
- Storage and transport temperature: -4ºF to 140ºF (-20ºC to 60ºC)
- Storage and transport humidity: 10–95% non-condensing
- Operating altitude: Sea level to 8,500’ (2,600 m)
- Supplemental oxygen: Recommended maximum supplemental oxygen flow: 4 L/min (VAuto mode); 15 L/min (CPAP, S, ST and T modes)
- Electromagnetic compatibility: Product complies with all applicable electromagnetic compatibility requirements (EMC) according to IEC60601-1-2, for residential, commercial, and light industry environments
- Air filter: Two-layered, powder-bonded, polyester non-woven fiber
- Air tubing: Flexible plastic, 1 x 6’6” (2 m)
- Air outlet: The 22 mm conical air outlet complies with ISO 5356-1
- IEC 60601-1 classification Class II (double insulation), Type CF
Note: The manufacturer reserves the right to change these specifications without notice.
Use My Insurance
Insurance Information Form
Complete your insurance information below so that we may qualify you. Allow one business day for processing.
Secure: This link is secure. Information will not be used for any other purpose.
Get Started Now
It only takes seconds…
You could save hundreds and
Free Shipping!
By submitting this information, I authorize Americare Respiratory Services, Inc., affiliates, subsidiaries, or parent company to contact me by phone. Rest assured that we value and protect your privacy and do not share your information with third parties.